Safety Incidents OSHA Severe Injury Reports · 2015–2025
2,004,209Inspections Most recent open 2026-07-13 Last loaded 2026-07-17

OSHA Inspection: FOOD & DRUG ADMINISTRATION

Planned inspection · Safety discipline

On , OSHA opened a planned safety inspection of FOOD & DRUG ADMINISTRATION in PACIFIC REGIONAL SOUTHWEST LABORATORY 19701 FAIRCHILD, IRVINE, CA 92612 (NAICS 541380). OSHA activity number 338983224.

Watch Food & Drug Administration — free Get an email when a new federal OSHA severe-injury report for Food & Drug Administration is published. One employer, no account, unsubscribe in one click.
Site address
PACIFIC REGIONAL SOUTHWEST LABORATORY 19701 FAIRCHILD
City
IRVINE
State
CA
ZIP
92612
Mailing
19701 FAIRCHILD, IRVINE, CA 92612
Inspection type
Planned (H)
Scope
Complete (A)
Discipline
Safety
Advance notice
No
Union status
A
Opened
Closing conference
Case closed
Last modified
Data loaded
NAICS code
541380
Employees
200
Ownership type
D

12 citations on file for this inspection.

1910.22 A02

Serious Gravity 5 1 instance 1 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.22(a)(2): Floor(s) of workroom(s) were not maintained in a clean and, so far as possible, a dry condition:   Microbiology Prep Area, Walk-In Freezer - On or about 03/27/2013, ice buildup posed slip hazards.
Recent events (1)
  • — Z (S) $0

1910.36 G02

Serious Gravity 5 1 instance 2 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.36(g)(2): Exit access(es) were not at least 28 inches (71.1 cm) wide at all points.   Control Substance Room exit access - On or about 3/28/2013, the exit access was less than 28 inches.
Recent events (1)
  • — Z (S) $0

1910.37 A03

Serious Gravity 5 2 instances 93 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.37(a)(3): Exit route(s) were not kept free and unobstructed:   a) Second floor, hallway adjacent to emergency exit door - On or about 3/28/2013, an emergency exit route was obstructed by pallets, boxes, and shelves with material.   b) Microbiology suite, hallway - On or about 3/28/2013, an emergency exit route was obstructed by pallets, boxes, and shelves with material.
Recent events (1)
  • — Z (S) $0

1910.101 B

Serious Gravity 5 1 instance 6 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.101(b): The in-plant handling, storage, and utilization of all compressed gases in cylinders, portable tanks, rail tankcars, or motor vehicle cargo tanks were not in accordance with Compressed Gas Association Pamphlet P-1-1965, which is incorporated by reference as specified in CFR 1910.6:   Mass spectrometry area - On or about 03/27/2013, a nitrogen tank was not secured from falling.
Recent events (1)
  • — Z (S) $0

1910.165 E

Serious Gravity 5 1 instance 1 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.165(e): Manually operated actuation devices for use in conjunction with employee alarms were not unobstructed, conspicuous and readily accessible:   Flammable waste storage - On or about 3/28/2013, a fire alarm manual pull box was obstructed by hazardous waste drums.
Recent events (1)
  • — Z (S) $0

1910.176 B

Serious Gravity 1 1 instance 7 exposed
Issued
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.176(b): Storage of material created a hazard:   Food chemistry area - On or about 3/27/2013, a broken glass receptacle was stored over the edge of a lab bench and could be knocked over.
Recent events (1)
  • — Z (S) $0

1910.176 C

Serious Gravity 5 2 instances 6 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.176(c): Storage areas were not kept free from accumulation of materials that constituted hazards from tripping, fire, explosion or pest harborage:  a) Mass spectrometry area - On or about 03/27/2013, catch containers were stored on the floor posing tripping hazards.  b) Drug chemistry main lab - On or about 03/28/2013, catch containers were stored on the floor posing tripping hazards.
Recent events (1)
  • — Z (S) $0

1910.215 A04

Serious Gravity 5 1 instance 1 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.215(a)(4): Work rest(s) on grinding machinery were not adjusted closely to the wheel with a maximum opening of one eighth inch:   Storage room - On or about 3/28/2013, the opening between the right wheel and the work rest was greater than one eighth inch.
Recent events (1)
  • — Z (S) $0

1910.215 B09

Serious Gravity 5 1 instance 1 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.215(b)(9): The distance between the abrasive wheel periphery(s) and the adjustable tongue or the end of the safety guard peripheral member at the top exceeded one fourth inch:   Storage room - On or about 3/28/2013, the distance between the right wheel and the end of the safety guard was greater than one fourth inch. The distance between the left wheel and the end of the safety guard was greater than one fourth inch.
Recent events (1)
  • — Z (S) $0

1910.303 B01 II

Serious Gravity 5 1 instance 8 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.303(b)(1)(ii): Electrical equipment was not free from recognized hazards, based on the mechanical strength and durability, including, for parts designed to enclose and protect other equipment, the adequacy of the protection thus provided:  Entomology wing, room 1105, Lab Hood FH03 - On or about 3/27/2013, an electrical outlet was inadequately protected by a loose faceplate.
Recent events (1)
  • — Z (S) $0

1910.303 B06

Serious Gravity 5 1 instance 8 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.303(b)(6): Unless identified for use in the operating environment, no conductors or equipment shall be located in damp or wet locations; where exposed to gases, fumes, vapors, liquids, or other agents that have a deteriorating effect on the conductors or equipment; or where exposed to excessive temperatures:   a) Entomology wing, Room 1105, Lab Hood FH05, acid storage - On or about 3/27/2013, electrical outlets not identified for appropriate use were located in an environment exposed to corrosive vapors.   b)  Entomology wing, Room 1105, Lab Hood FH03 - On or about 3/27/2013, electrical outlets not identified for appropriate use were located in an environment exposed to corrosive vapors.
Recent events (1)
  • — Z (S) $0

1910.305 G01 IV A

Serious Gravity 5 1 instance 1 exposed
Issued
Abate by
Penalty
Initial $0.00 · Current $0.00
29 CFR 1910.305(g)(1)(iv)(A): Flexible cords and/or cables were used as a substitute for the fixed wiring of a structure:   PFGE room - On or about 3/27/2013, an extension cord was used as permanent wiring to supply power to a refrigerator.
Recent events (1)
  • — Z (S) $0

View FOOD & DRUG ADMINISTRATION's full OSHA safety record →