OSHA Inspection: COMMAND MEDICAL PRODUCTS, INC.

Complaint inspection · Safety discipline

On , OSHA opened a complaint safety inspection of COMMAND MEDICAL PRODUCTS, INC. in 15 SIGNAL AVENUE, ORMOND BEACH, FL 32174 (NAICS 326111). OSHA activity number 339163347.

What this inspection record means

OSHA opens inspections for many reasons — routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.

Watch Command Medical Products, INC. — free Get an email when a new federal OSHA severe-injury report for Command Medical Products, INC. is published. One employer, no account, unsubscribe in one click.
Site address
15 SIGNAL AVENUE
City
ORMOND BEACH
State
FL
ZIP
32174
Mailing
15 SIGNAL AVE, ORMOND BEACH, FL 32174
Inspection type
Complaint (B)
Scope
Complete (A)
Discipline
Safety
Advance notice
No
Union status
B
Opened
Closing conference
Case closed
Last modified
Data loaded
NAICS code
326111
Employees
56
Ownership type
A

2 citations on file for this inspection.

1910.305 B01

Serious Gravity 5 1 instance 40 exposed
Issued
Sep 27, 2013
Penalty
Initial $2,380 · Current $2,380
29 CFR 1910.305(b)(1): Conductors entering boxes, cabinets, or fittings were not protected from abrasions:  a.  On or about June 27, 2013, in the production department, the electrical panel on the RF-14 sealing machine had energized wires running through the door which was closed with duct tape, exposing employees to an electrical hazard.
Recent events (2)
  • — I (S) $2380
  • — Z (S) $2380

1910.305 G02 III

Other-than-serious 1 instance 40 exposed
Issued
Sep 27, 2013
Penalty
Initial $0 · Current $0
29 CFR 1910.305(g)(2)(iii): Flexible cords and cables were not connected to devices and fittings so that strain relief was provided that would prevent pull from being directly transmitted to joints or terminal screws:  a. On or about June 27, 2013, in the production department near the RF-11 sealing machine, the electrical conduit that was hanging from the ceiling did not have adequate strain relief.
Recent events (2)
  • — I (O) $0
  • — Z (O) $0

View Command Medical Products, INC.'s full OSHA safety record →

This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 339163347.

Look up any company's OSHA accident reports by company, or browse severe injury reports by year, state, and company.