ARVADA, CO —
OSHA Inspection: SORIN GROUP USA, INC.
Planned inspection · Health discipline
At a glance
On , OSHA opened a planned health inspection of SORIN GROUP USA, INC. in 14401 W. 65TH WAY, ARVADA, CO 80004 (NAICS 339112). OSHA activity number 339481756.
OSHA opens inspections for many reasons — routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.
Where did this inspection happen?
- Establishment
- SORIN GROUP USA, INC.
- Site address
- 14401 W. 65TH WAY
- City
- ARVADA
- State
- CO
- ZIP
- 80004
- Mailing
- 14401 W. 65TH WAY, ARVADA, CO 80004
What kind of inspection was it?
- Inspection type
- Planned (H)
- Scope
- Complete (A)
- Discipline
- Health
- Advance notice
- No
- Union status
- B
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 339112
- Employees
- 250
- Ownership type
- A
Citations
3 citations on file for this inspection.
1910.23 C01
- Issued
- Mar 14, 2014
- Penalty
- Initial $1,913 · Current $1,913
General-duty citation text
29 CFR 1910.23(c)(1): Open-sided floors and/or platforms four feet or more above adjacent floor or ground level were not guarded with standard railings (or equivalent) and toeboards: a) On or about November 12, 2013, and at times prior, employees were exposed to fall hazards of greater than 4 feet when they worked in the storage area above the maintenance office because the mezzanine level floor was not equipped with standard railings on all sides.
Recent events (2)
- — I (O) $1913
- — Z (S) $1913
1910.1052 D02
- Issued
- Mar 14, 2014
- Penalty
- Initial $2,250 · Current $2,250
General-duty citation text
29 CFR 1910.1052(d)(2): The employer did not perform initial monitoring to determine each employee's exposure to methylene chloride: a) On or about November 12, 2013, and at times prior, employees were exposed to unknown concentrations of methylene chloride because the employer failed to conduct an initial exposure assessment for employees of the Ortho fuser line that used a solvent mixture, which contained 50% methylene chloride, as an adhesive for assembly of medical parts.
Recent events (2)
- — I (O) $2250
- — Z (S) $2250
1910.305 G01 IV
- Issued
- Mar 14, 2014
- Abate by
- Mar 28, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.305(g)(1)(iv): Flexibles cords and cables were used for purposes prohibited by subparagraphs (a) through (e) of this paragraph: a) On or about November 12, 2013, and at times prior, employees were exposed to electrical shock and fire hazards because flexible cords were attached to building surfaces and used as a substitute for fixed wiring (such as conduit) in the maintenance shop and the CVE room. Flexible cords were attached to the wall in the maintenance shop and attached to the ceiling and direct wired into standard metal outlet boxes in the CVE room.
Recent events (2)
- — I (O) $0
- — Z (O) $0
More inspections at Sorin Group USA, INC.
View Sorin Group USA, INC.'s full OSHA safety record →
More inspections in this industry (NAICS 339112)
More inspections in CO
Source
This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 339481756.