SAN PEDRO, CA ·
OSHA Inspection: U.S. CUSTOMS AND BORDER PROTECTION - LOS ANGELES LABORATORY
Planned inspection · Safety discipline
At a glance
On , OSHA opened a planned safety inspection of U.S. CUSTOMS AND BORDER PROTECTION - LOS ANGELES LABORATORY in 300 S FERRY ST SUITE 1300, SAN PEDRO, CA 90731 (NAICS 922120). OSHA activity number 339488538.
OSHA opens inspections for many reasons: routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.
Where did this inspection happen?
- Site address
- 300 S FERRY ST SUITE 1300
- City
- SAN PEDRO
- State
- CA
- ZIP
- 90731
- Mailing
- 300 S FERRY ST SUITE 1300, SAN PEDRO, CA 90731
What kind of inspection was it?
- Inspection type
- Planned (H)
- Scope
- Complete (A)
- Discipline
- Safety
- Advance notice
- No
- Union status
- Union (A)
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 922120
- Employees
- 27
- Ownership type
- Federal government (D)
Citations
8 citations on file for this inspection.
1910.157 C01
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.157(c)(1): Portable fire extinguishers were not mounted, located and identified so that they were readily accessible without subjecting the employees to injuries: Rm 1330C - On or about 11/14/13, access to a fire extinguisher was blocked by a ladder and chair.
Recent events (1)
- · Z (S) $0
1910.157 E02
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.157(e)(2): Portable fire extinguishers were not visually inspected at least monthly: a. Library - On or about 11/14/13, a fire extinguisher was not inspected at least monthly. b. Rm 1329 - On or about 11/14/13, a fire extinguisher was not inspected at least monthly. c. Rm 1330C - On or about 11/14/13, a fire extinguisher was not inspected at least monthly.
Recent events (1)
- · Z (S) $0
1910.176 C
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.176(c): Storage areas were not kept free from accumulation of materials that constituted hazards from tripping, fire, explosion or pest harborage: a. Doors 1317-1319 - On or about 11/14/13, items were stored on the floor and posed tripping hazards. b. Custodial closet - On or about 11/14/13, items in the aisle posed tripping hazards. c. Rm 1353, hood H1 flammable locker - On or about 11/14/13, a strong oxidizer was stored with flammables and constituted a fire hazard. d. Rm 1353, aisle - On or about 11/14/13, a catch container posed a tripping hazard. e. Rm 1352, aisle - On or about 11/14/13, a catch container posed a tripping hazard. f. Rm 1307 - On or about 11/14/13, combustibles were stored within 3 feet of a personal heater.
Recent events (1)
- · Z (S) $0
1910.215 A04
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.215(a)(4): Work rest(s) on grinding machinery were not adjusted closely to the wheel with a maximum opening of one eighth inch: Rm 1329 - On or about 3/28/2013, the opening between the grinding wheels and the work rests were greater than one eighth inch.
Recent events (1)
- · Z (S) $0
1910.303 B02
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.303(b)(2): Listed or labeled electrical equipment was not used or installed in accordance with instructions included in the listing or labeling: a. Front office - On or about 11/14/13, an extension cord was daisy-chained to a surge protector. b. Front office, student desk area - On or about 11/14/13, two surge protectors were daisy-chained together. c. Library - On or about 11/14/13, an extension cord was daisy-chained to a surge protector. d. Rm 1303 - On or about 11/14/13, two surge protectors were daisy-chained together.
Recent events (1)
- · Z (S) $0
1910.305 G01 IV A
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.305(g)(1)(iv)(A): Flexible cords and/or cables were used as a substitute for the fixed wiring of a structure: a. Front office - On or about 11/14/13, an extension cord was used as permanent wiring. b. Library - On or about 11/14/13, extension cords were used as permanent wiring.
Recent events (1)
- · Z (S) $0
1910.1200 F05 I
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1200(f)(5)(i): The employer did not ensure that each container of hazardous chemicals in the workplace was labeled, tagged or marked with the identity of the hazardous chemical(s) contained therein: Rm 1356, hood H17 - On or about 11/14/13, a spray bottle was not labeled with its chemical contents.
Recent events (1)
- · Z (S) $0
1910.1200 F09
- Issued
- Dec 20, 2013
- Abate by
- Jan 17, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1200(f)(9): The employer did not ensure that labels or other forms of warning were legible, in English, and prominently displayed on the container, or readily available in the work area throughout each work shift: Rm 1356, hood H17 - On or about 11/14/13, chemical bottle labels were illegible.