DILLINGHAM, AK —
OSHA Inspection: BRISTOL BAY AREA HEALTH CORPORATION
Planned inspection · Health discipline
At a glance
On , OSHA opened a planned health inspection of BRISTOL BAY AREA HEALTH CORPORATION in 6000 KANAKANAK ROAD, DILLINGHAM, AK 99576 (NAICS 622110). OSHA activity number 340749639.
Where did this inspection happen?
- Establishment
- BRISTOL BAY AREA HEALTH CORPORATION
- Site address
- 6000 KANAKANAK ROAD
- City
- DILLINGHAM
- State
- AK
- ZIP
- 99576
- Mailing
- PO BOX 130, DILLINGHAM, AK 99576
What kind of inspection was it?
- Inspection type
- Planned (H)
- Scope
- Complete (A)
- Discipline
- Health
- Advance notice
- No
- Union status
- B
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 622110
- Employees
- 360
- Ownership type
- A
Citations
11 citations on file for this inspection.
1910.134 C02 II
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $4,250 · Current $2,125 Reduced
General-duty citation text
29 CFR 1910.134(c)(2)(ii): The employer did not establish and implement those elements of a written program necessary to ensure that any employee using a respirator voluntarily was medically able to use that respirator, and that the respirator was cleaned, stored, and maintained so that its use does not present a health hazard to the user: Boiler Plant, Biohazard Autoclave: On or about June 30, 2015 and at times prior thereto, the employer did not implement medical surveillance or ensure respirators were properly cleaned, stored, and maintained for employees using 3M and AO Safety, half-mask, negative pressure respirators. This condition placed a physiological burden for voluntary respirator users and exposed users to skin contact hazards when the respirator was donned.
Recent events (2)
- — I (S) $2125
- — Z (S) $4250
1910.213 H04
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $3,400 · Current $1,700 Reduced
General-duty citation text
29 CFR 1910.213(h)(4): Radial saw(s) were not installed in a manner so as to cause the cutting head to return gently to the starting position when released by the operator: Woodshop: On or about June 30, 2015 and at times prior thereto, the Delta Radial Arm Saw did not retract to its starting position when released. This condition exposed employees to amputation hazards.
Recent events (2)
- — I (S) $1700
- — Z (S) $3400
1910.304 G05
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $4,250 · Current $2,125 Reduced
General-duty citation text
29 CFR 1910.304(g)(5): The path to ground from circuits, equipment, and enclosures was not permanent, continuous, and effective: Boiler Plant, Shop Bench: On or about June 30, 2015 and at times prior thereto, the power strip with outlets labeled L1-2 and L1-4 indicated false ground (FG). This condition exposed employee to employees to electrical shock hazards from the 20 amp, 120 volt service.
Recent events (2)
- — I (S) $2125
- — Z (S) $4250
1910.1030 H05 I A
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $4,250 · Current $2,125 Reduced
General-duty citation text
29 CFR 1910.1030(h)(5)(i)(A): The employer did not maintain a sharps injury log that contained at a minimum the type and brand of device involved in the incident: Kanakanak Hospital: On or about July 1, 2015 and at times prior thereto, the employer's sharps injury log did not contain the type and brand of device associated with the sharps injury. This condition exposed employees to additional sharps injuries by not fully evaluating the cause of prior sharps injuries.
Recent events (2)
- — I (S) $2125
- — Z (S) $4250
1910.1030 C01 IV
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1030(c)(1)(iv): The Exposure Control Plan was not reviewed and updated at least annually: Kanakanak Hospital: On or about July 1, 2015 and at times prior thereto, the employer's Exposure Control Plan had not been reviewed and updated annually. This condition exposed employees to additional sharps injuries by not fully evaluating new methods of eliminating bloodborne pathogen hazards in the workplace.
Recent events (2)
- — I (S) $0
- — Z (S) $0
1910.1030 C01 IV B
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1030(c)(1)(iv)(B): The review and update of the exposure control plan did not document annually consideration and implementation of appropriate commercially available and effective safer medical devices designed to eliminate or minimize occupational exposure: Kanakanak Hospital, Laboratory: On or about June 30, 2015 and at times prior thereto, the employer's had not documented and reviewed replacements for glass blood bank tubes and glass Bactec blood culture bottles. This condition exposed employees to additional sharps injuries by not fully evaluating new methods of eliminating bloodborne pathogen hazards in the workplace.
Recent events (2)
- — I (S) $0
- — Z (S) $0
1910.1030 D02 XIV
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1030(d)(2)(xiv): Equipment which became contaminated with blood or other potentially infectious materials was not examined prior to servicing or shipping and was not decontaminated as necessary: Boiler Plant: On or about June 30, 2015 and at times prior thereto, the bottom surface of the interior of the Mark-Costello autoclave was contaminated with blood or other potentially infectious material. This condition exposed employee to contamination from contact with the substance.
Recent events (2)
- — I (S) $0
- — Z (S) $0
1910.37 B02
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.37(b)(2): Each exit was not clearly visible and marked by a sign reading "Exit": Kanakanak Hospital, Laboratory: On or about June 30, 2015 and at times prior thereto, exit(s) were not marked with sign(s) reading "Exit". This condition exposed employees to delay in accessing the exit in the event of an emergency.
Recent events (2)
- — I (O) $0
- — Z (O) $0
1910.101 B
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.101(b): The in-plant handling, storage, and utilization of all compressed gases in cylinders, portable tanks, rail tankcars, or motor vehicle cargo tanks were not in accordance with Compressed Gas Association Pamphlet P-1-1965, which is incorporated by reference as specified in CFR 1910.6: (a) Boiler Plant, Tool Storage Room: On or about June 30, 2015 and at times prior thereto, the storage room for the oxygen and acetylene cylinders was not well-ventilated. This condition exposed employees to smoke-inhalation hazards. (b) Boiler Plant, Tool Storage Room: On or about June 30, 2015 and at times prior thereto, oxygen and acetylene in storage were not separated by a minimum of 20 feet or fire-resistant partition. This condition exposed employees to smoke-inhalation hazards.
Recent events (2)
- — I (O) $0
- — Z (O) $0
1910.147 C04 I
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.147(c)(4)(i): Procedures were not developed, documented and utilized for the control of potentially hazardous energy when employees were engaged in activities covered by this section: (a) Sterile Supply: On or about June 30, 2015 and at times prior thereto, procedures for energy isolation of the Steris V-120 Prevac Steam Sterilizer had not been developed, documented, and utilized. This condition exposed employees to struck-by and contact hazards from electrical, steam, hot surfaces, and high pressure sources when servicing the equipment. (b) Building 401: On or about June 30, 2015 and at times prior thereto, procedures for energy isolation of the Kaeser Aircenter SM10 had not been developed, documented, and utilized. This condition exposed employees to struck-by and contact hazards from electrical and high pressure sources when servicing the equipment.
Recent events (2)
- — I (O) $0
- — Z (O) $0
1910.1200 F06 II
- Issued
- Jul 29, 2015
- Abate by
- Aug 26, 2015
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1200(f)(6)(ii): Except as provided in 29 CFR 1910.1200(f)(7) and 29 CFR 1910.1200(f)(8), the employer did not ensure that each container of hazardous chemicals in the workplace was labeled, tagged or marked with the product identifier and words, pictures, symbols, or combination thereof, which provide at least general information regarding the hazards of the chemicals and which, in conjunction with the other information immediately available to employees under the hazard communication program, would provide employees with the specific information regarding the physical and health hazards of the hazardous chemical: Boiler Plant Building 601: On or about June 30, 2015 and at times prior thereto, the stationary container for the 15/1 mix of autoclave deodorizer was not labeled. This condition limited employee information of the contents of the container and the use of proper precautions when handling the liquid.
Recent events (2)
- — I (O) $0
- — Z (O) $0
More inspections at Bristol BAY Area Health Corporation
View Bristol BAY Area Health Corporation's full OSHA safety record →
More inspections in this industry (NAICS 622110)
More inspections in AK
Source
This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 340749639.
Look up any company's OSHA accident reports by company, or browse severe injury reports by year, state, and company.