GOLDEN, CO ·
OSHA Inspection: WEATHERFORD LABORATORIES, INC.
Complaint inspection · Safety discipline
At a glance
On , OSHA opened a complaint safety inspection of WEATHERFORD LABORATORIES, INC. in 16161 TABLE MOUNTAIN PKWY, GOLDEN, CO 80403 (NAICS 213112). OSHA activity number 341444305.
OSHA opens inspections for many reasons: routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.
Where did this inspection happen?
- Establishment
- WEATHERFORD LABORATORIES, INC.
- Site address
- 16161 TABLE MOUNTAIN PKWY
- City
- GOLDEN
- State
- CO
- ZIP
- 80403
- Mailing
- 16161 TABLE MOUNTAIN PARKWAY, GOLDEN, CO 80403
What kind of inspection was it?
- Inspection type
- Complaint (B)
- Scope
- Partial (B)
- Discipline
- Safety
- Advance notice
- No
- Union status
- B
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 213112
- Employees
- 62
- Ownership type
- A
Citations
2 citations on file for this inspection.
1910.106 E06 I
- Issued
- Jun 2, 2016
- Abate by
- Jun 21, 2016
- Penalty
- Initial $2,800 · Current $1,680 Reduced
General-duty citation text
29 CFR 1910.106(e)(6)(i): Adequate precautions were not taken to prevent the ignition of flammable vapors. a) On or about April 29, 2016, and at times prior, employees were exposed to fire hazards while they dispensed flammable liquids including but not limited to, methanol and toluene from 55 gallon drums, which did not have grounding straps, into containers which were not bonded to the fill stem.
Recent events (2)
- · I (S) $1680
- · Z (S) $2800
1910.106 E06 II
- Issued
- Jun 2, 2016
- Abate by
- Jun 21, 2016
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.106(e)(6)(ii): Grounding. Category 1 or 2 flammable liquids, or Category 3 flammable liquids with a flashpoint below 100 °F (37.8 °C), were not dispensed into containers where the nozzle and container were electrically interconnected. b) On or about April 29, 2016, and at times prior, employees were exposed to fire hazards while they dispensed flammable liquids including but not limited to, methanol and toluene from 55 gallon drums into containers which were not bonded to the fill stem.
Recent events (2)
- · I (S) $0
- · Z (S) $0
More inspections at Weatherford Laboratories, INC.
View Weatherford Laboratories, INC.'s full OSHA safety record →
More inspections in this industry (NAICS 213112)
More inspections in CO
Source
This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 341444305.
Look up any company's OSHA accident reports by company, or browse severe injury reports by year, state, and company.