SILVERDALE, WA ·
OSHA Inspection: NAVY BRANCH HEALTH CLINIC BANGOR
Planned inspection · Safety discipline
At a glance
On , OSHA opened a planned safety inspection of NAVY BRANCH HEALTH CLINIC BANGOR in 2050 BARB STREET SUITE A, SILVERDALE, WA 98315 (NAICS 622110). OSHA activity number 343811543.
OSHA opens inspections for many reasons: routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.
Where did this inspection happen?
- Establishment
- NAVY BRANCH HEALTH CLINIC BANGOR
- Site address
- 2050 BARB STREET SUITE A
- City
- SILVERDALE
- State
- WA
- ZIP
- 98315
- Mailing
- 2050 BARB STREET SUITE A, SILVERDALE, WA 98315
What kind of inspection was it?
- Inspection type
- Planned (H)
- Scope
- Complete (A)
- Discipline
- Safety
- Advance notice
- No
- Union status
- Non-union (B)
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 622110
- Employees
- 16
- Ownership type
- Federal government (D)
Citations
2 citations on file for this inspection.
1910.303 B02
- Issued
- Mar 12, 2019
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.303(b)(2): Listed or labeled electrical equipment was not used or installed in accordance with instructions included in the listing or labeling: a) Pharmacy Room H1009; A Keurig, Magic Chef microwave and Sanyo refrigerator were plugged into a surge protector power strip (RPT). The UL listing states that relocatable power taps (RPTs) are designed for low amperage computer and audiovisual types of equipment, and not appliances. This exposes employees to possible electrical fire hazards.
Recent events (1)
- · Z (S) $0
1910.37 B05
- Issued
- Mar 12, 2019
- Abate by
- Apr 5, 2019
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.37(b)(5): Each doorway or passage along an exit access that could be mistaken for an exit was not marked "Not an Exit" or similar designation, or be identified by a sign indicating its actual use (e.g. closet): a) Exam Room B-1013. A locked door was not marked NOT AN EXIT in an exam room with two doors