Safety Incidents OSHA Severe Injury Reports · 2015–2025
5,198,547Inspections Most recent open 2026-09-01 Last loaded 2026-09-04

OSHA Inspection: NEXTPHASE MEDICAL DEVICES LLC

Complaint inspection · Health discipline

On , OSHA opened a complaint health inspection of NEXTPHASE MEDICAL DEVICES LLC in 88 AIRPORT DRIVE, ROCHESTER, NH 03867 (NAICS 339112). OSHA activity number 344194253.

What this inspection record means

OSHA opens inspections for many reasons: routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.

Watch Nextphase Medical Devices LLC for free Get an email when a new federal OSHA severe-injury report for Nextphase Medical Devices LLC is published. One employer, no account, unsubscribe in one click.
Site address
88 AIRPORT DRIVE
City
ROCHESTER
State
NH
ZIP
03867
Mailing
88 AIRPORT DRIVE, ROCHESTER, NH 03867
Inspection type
Complaint (B)
Scope
Partial (B)
Discipline
Health
Advance notice
No
Union status
Non-union (B)
Opened
Closing conference
Case closed
Last modified
Data loaded
NAICS code
339112
Employees
180
Ownership type
Private (A)

1 citation on file for this inspection.

1910.303 B02

Serious Gravity 5 1 instance 2 exposed
Issued
Nov 20, 2019
Abate by
Mar 1, 2020
Penalty
Initial $8,525 · Current $4,263 Reduced
29 CFR 1910.303(b)(2): Listed or labeled electrical equipment was not used or installed in accordance with instructions included in the listing or labeling:    Location: pad print cleaning area    Instance 1:   On or about 7/31/19, A Nutone Allure WS1 range hood designed for domestic cooking and not to be used to exhaust hazardous or explosive materials and vapors was used to ventilate an area where a flammable liquid, acetone, was used to clean parts.     Instance 2:  On or about 7/31/19, a relocatable power tap UL listed for use with low power electronic devices was use to power two range hood fan motors.
Recent events (2)
  • · I (S) $4262.5
  • · Z (S) $8525

This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). OSHA publishes its own view of this case as inspection number 344194253.

Look up any company's OSHA accident reports by company, or browse severe injury reports by year, state, and company.