Safety Incidents OSHA Severe Injury Reports · 2015–2025
5,189,992Inspections Most recent open 2026-07-18 Last loaded 2026-07-22

OSHA Inspection: NOAA MONTLAKE LABORATORY- NW FISHERIES SCIENCE CENTER

Planned inspection · Health discipline

On , OSHA opened a planned health inspection of NOAA MONTLAKE LABORATORY- NW FISHERIES SCIENCE CENTER in 2725 MONTLAKE BOULEVARD EAST, SEATTLE, WA 98112 (NAICS 541715). OSHA activity number 345803027.

Watch Noaa Montlake Laboratory- NW Fisheries Science Center — free Get an email when a new federal OSHA severe-injury report for Noaa Montlake Laboratory- NW Fisheries Science Center is published. One employer, no account, unsubscribe in one click.
Site address
2725 MONTLAKE BOULEVARD EAST
City
SEATTLE
State
WA
ZIP
98112
Mailing
2725 MONTLAKE BOULEVARD EAST, SEATTLE, WA 98112
Inspection type
Planned (H)
Scope
Complete (A)
Discipline
Health
Advance notice
No
Union status
B
Opened
Closing conference
Case closed
Last modified
Data loaded
NAICS code
541715
Employees
300
Ownership type
D

3 citations on file for this inspection.

1910.215 A04

Repeat Gravity 5 1 instance 1 exposed
Issued
Mar 17, 2022
Penalty
Initial $0 · Current $0
29 CFR  1910.215(a)(4):Work rest(s) on grinding machinery were not adjusted closely to the wheel with a maximum opening of one eighth inch:  (a)  East Building, Maintenance work area adjacent to boiler room;  A grinder (brand name and model number were covered with employee labels) had the left side grinding stone located ? inch from the tool rest.  NOAA NW Fisheries was previously cited for a violation of this occupational safety and health standard, which was contained in OSHA inspection number 1521897, citation number 2, item number 3 and was affirmed as a final order on 01/12/2022, with respect to a workplace located at Port Orchard, WA.
Recent events (1)
  • — Z (R) $0

1910.1200 F06 II

Repeat 4 instances 4 exposed
Issued
Mar 17, 2022
Penalty
Initial $0 · Current $0
29 CFR  1910.1200(f)(6)(ii): Except as provided in 29 CFR 1910.1200(f)(7) and 29 CFR 1910.1200(f)(8), the employer did not ensure that each container of hazardous chemicals in the workplace was labeled, tagged or marked with the product identifier and words, pictures, symbols, or combination thereof, which provide at least general information regarding the hazards of the chemicals and which, in conjunction with the other information immediately available to employees under the hazard communication program, would provide employees with the specific information regarding the physical and health hazards of the hazardous chemical:  (a)	East Building Laboratory- Room 456 E, Physiology Lab;  A secondary bottle of 75% ethanol did not have hazard label warnings.   (b)	East Building Laboratory- Room 433E, Marine Microbes and Toxins Lab;  A secondary bottle of 100% isopropanol did not have hazard label warnings.   (c)	East Building Laboratory- Room 354E, Ecotoxicology Lab;  A secondary bottle of 70% ethanol did not have hazard label warnings.   (d)	East Building Laboratory- Room 210E, Watershed Lab;  A secondary bottle of 70% ethanol did not have hazard label warnings.   NOAA NW Fisheries was previously cited for a violation of this occupational safety and health standard, which was contained in OSHA inspection number 1546470, citation number 1, item number 1 and was affirmed as a final order on 10/18/2021, with respect to a workplace located at Seattle, WA.  NOAA NW Fisheries was previously cited for a violation of this occupational safety and health standard, which was contained in OSHA inspection number 1521897, citation number 2, item number 3 and was affirmed as a final order on 01/12/2022, with respect to a workplace located at Port Orchard, WA.
Recent events (1)
  • — Z (R) $0

1910.303 G01 II

Other-than-serious 1 instance 2 exposed
Issued
Mar 17, 2022
Penalty
Initial $0 · Current $0
29 CFR  1910.303(g)(1)(ii):The required working space about electric equipment rated 600 volts, nominal, or less to ground, was used for storage:  a)	 East Building Laboratory- Third floor in between Room 345 and 343E;  Several boxes, plastic storage containers and empty 5 gallon plastic buckets were stored underneath breaker panel no 3J.
Recent events (1)
  • — Z (O) $0

This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 345803027.

Look up any company's OSHA accident reports by company, or browse severe injury reports by year, state, and company.