KANSAS CITY, MO ·
OSHA Inspection: SCD PROBIOTICS
Planned inspection · Health discipline
At a glance
On , OSHA opened a planned health inspection of SCD PROBIOTICS in 1710 WALNUT STREET, KANSAS CITY, MO 64108 (NAICS 311999). OSHA activity number 347235418.
OSHA opens inspections for many reasons: routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.
Where did this inspection happen?
- Establishment
- SCD PROBIOTICS
- Site address
- 1710 WALNUT STREET
- City
- KANSAS CITY
- State
- MO
- ZIP
- 64108
- Mailing
- 1710 WALNUT STREET, KANSAS CITY, MO 64108
What kind of inspection was it?
- Inspection type
- Planned (H)
- Scope
- Complete (A)
- Discipline
- Health
- Advance notice
- No
- Union status
- Non-union (B)
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 311999
- Employees
- 8
- Ownership type
- Private (A)
Citations
3 citations on file for this inspection.
1910.303 B02
- Issued
- May 22, 2024
- Abate by
- Jun 18, 2024
- Penalty
- Initial $3,687 · Current $2,212 Reduced
General-duty citation text
29 CFR 1910.303(b)(2): Listed or labeled electrical equipment was not used or installed in accordance with instructions included in the listing or labeling: At the worksite located at 1710 Walnut Street Kansas City, MO: Employees were exposed to shock and fire hazards in multiple areas of the lab. Laboratory equipment including but not limited to incubators, blenders, PCR prep stations, and fridges were plugged into power taps that were not listed and labeled for the level of power. a) A PCR prep station plugged into a power tap in the PCR room, as well as another active power tap located on the same table b) A white power tap plugged into the wall of the chemistry room c) Second white power tap plugged into an extension cord in the chemistry room d) A white power tap in the main lab area with an incubator plugged in e) A black power tap in the main lab area f) A second white power tab in the main lab area g) A white power tap by the hoods and washing machine h) A white power tap used in the back storage room
Recent events (3)
- · F (S) $2212.2
- · C (S) $3687
- · Z (S) $3687
1910.305 G01 IV A
- Issued
- May 22, 2024
- Abate by
- Jun 18, 2024
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.305 (g)(1)(iv)(A): Flexible cords and/or cables were used as a substitute for the fixed wiring of a structure: At the worksite located at 1710 Walnut Street Kansas City, MO: Employees were exposed to shock, burn, and or electrocution hazards in the back storage room and other lab rooms. Temporary cords were used in lieu of fixed wiring to power the refrigerator and in the use of power taps throughout the lab. a) A PCR prep station plugged into a power tap in the PCR room, as well as another active power tap located on the same table b) A white power tap plugged into the wall of the chemistry room c) Second white power tap plugged into an extension cord in the chemistry room d) A white power tap in the main lab area with an incubator plugged in e) A black power tap in the main lab area f) A second white power tab in the main lab area g) A white power tap by the hoods and washing machine h) A white power tap used in the back storage room
Recent events (3)
- · F (S) $0
- · C (S) $0
- · Z (S) $0
1910.1450 E01
- Issued
- May 22, 2024
- Abate by
- Jun 18, 2024
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.1450(e)(1): Where hazardous chemicals were used in the workplace, the employer did not develop and carry out the provisions of a written Chemical Hygiene Plan which was capable of protecting employees from health hazards associated with hazardous chemicals: At the worksite located at 1710 Walnut Street Kansas City, MO: Employees were exposed to chemical hazards. Employees worked in the lab without a written Chemical Hygiene Plan to protect employees from hazardous chemicals such as, but not limited to, acetonitrile, paraquat, pyridine, and toluene.
Recent events (3)
- · F (O) $0
- · C (O) $0
- · Z (O) $0