RENO, NV —
OSHA Inspection: AGRICULTURE RESEARCH SERVICES
Planned inspection · Safety discipline
At a glance
On , OSHA opened a planned safety inspection of AGRICULTURE RESEARCH SERVICES in 920 VALLEY ROAD, RENO, NV 89512 (NAICS 541720). OSHA activity number 339896888.
OSHA opens inspections for many reasons — routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.
Where did this inspection happen?
- Establishment
- AGRICULTURE RESEARCH SERVICES
- Site address
- 920 VALLEY ROAD
- City
- RENO
- State
- NV
- ZIP
- 89512
- Mailing
- 920 VALLEY ROAD, RENO, NV 89512
What kind of inspection was it?
- Inspection type
- Planned (H)
- Scope
- Complete (A)
- Discipline
- Safety
- Advance notice
- No
- Union status
- B
When did the case open and close?
- Opened
- Closing conference
- Case closed
- Last modified
- Data loaded
Establishment context
- NAICS code
- 541720
- Employees
- 17
- Ownership type
- D
Citations
7 citations on file for this inspection.
1910.37 B02
- Issued
- Sep 22, 2014
- Abate by
- Oct 16, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.37(b)(2): Each exit was not clearly visible and marked by a sign reading "Exit": (a) At the USDA Green House and Laboratory Room 28A, doors in the exit pathway were not marked or labeled "EXIT".
Recent events (1)
- — Z (S) $0
1910.37 B05
- Issued
- Sep 22, 2014
- Abate by
- Oct 16, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.37(b)(5): Each doorway or passage along an exit access that could be mistaken for an exit was not marked "Not an Exit" or similar designation, or be identified by a sign indicating its actual use (e.g. closet): (a) At the USDA DNA Lab, the door leading to room 23 which could be mistaken for an exit was not labeled or marked "NOT EXIT".
Recent events (1)
- — Z (S) $0
1910.303 B01
- Issued
- Sep 22, 2014
- Abate by
- Oct 16, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.303(b)(1): 29 CFR 1910.303(b)(1): Electrical equipment was not free from recognized hazards that were likely to cause death or serious physical harm to employees. (a) At the USDA Entomology, Modular Lab room 104, the energized duplex convenience outlet receptacle did not have a cover.
Recent events (1)
- — Z (S) $0
1910.304 B03 I
- Issued
- Sep 22, 2014
- Abate by
- Oct 16, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.304(b)(3)(i): All 125-volt, single-phase, 15- and 20-ampere receptacles installed in wet locations were not equipped with ground-fault circuit-interrupter protection for personnel. (a) At the USDA male/female rest rooms, employee break room and green room, energized wall outlet receptacles within three feet distance to the sink did not have ground fault circuit interrupter protection.
Recent events (1)
- — Z (S) $0
1910.36 G02
- Issued
- Sep 22, 2014
- Abate by
- Oct 16, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.36(g)(2): Exit access(es) were not at least 28 inches (71.1 cm) wide at all points. (a) At the USDA room 22, the only exit/egress for the employee was narrowed down to twenty four inches and obstructed by a wooden file cabinet and desk.
Recent events (1)
- — Z (O) $0
1910.157 C01
- Issued
- Sep 22, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.157(c)(1): Portable fire extinguishers were not mounted, located and identified so that they were readily accessible without subjecting the employees to injuries. (a) At the USDA Entomology, Modular Lab room 104, the portable fire extinguisher was obstructed by broom, dustpan and other cleaning tools.
Recent events (1)
- — Z (O) $0
1910.303 B02
- Issued
- Sep 22, 2014
- Abate by
- Oct 16, 2014
- Penalty
- Initial $0 · Current $0
General-duty citation text
29 CFR 1910.303(b)(2): Listed or labeled electrical equipment was not used or installed in accordance with instructions included in the listing or labeling. (a) At the USDA DNA Lab room 21, the energized re-locatable power tap was used next a sink to power the laboratory equipment.
Recent events (1)
- — Z (O) $0
More inspections at Agriculture Research Services
View Agriculture Research Services's full OSHA safety record →
More inspections in this industry (NAICS 541720)
More inspections in NV
Source
This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 339896888.