Safety Incidents OSHA Severe Injury Reports · 2015–2025
5,189,992Inspections Most recent open 2026-07-18 Last loaded 2026-07-22

OSHA Inspection: CARESTREAM HEALTH, INC.

Referral inspection · Safety discipline

On , OSHA opened a referral safety inspection of CARESTREAM HEALTH, INC. in 2000 HOWARD SMITH AVENUE WEST, WINDSOR, CO 80550 (NAICS 423450). OSHA activity number 346268139.

What this inspection record means

OSHA opens inspections for many reasons — routine scheduling under a national or local emphasis program, an employee complaint or referral, or a follow-up after a reported injury. Opening or conducting an inspection is not itself an allegation or a finding that this employer broke any rule; any findings appear as the citations listed below, and citations can be contested, reduced, or withdrawn.

Watch Carestream Health, INC. — free Get an email when a new federal OSHA severe-injury report for Carestream Health, INC. is published. One employer, no account, unsubscribe in one click.
Site address
2000 HOWARD SMITH AVENUE WEST
City
WINDSOR
State
CO
ZIP
80550
Mailing
2000 HOWARD SMITH AVENUE WEST, WINDSOR, CO 80550
Inspection type
Referral (C)
Scope
Partial (B)
Discipline
Safety
Advance notice
No
Union status
B
Opened
Closing conference
Case closed
Last modified
Data loaded
NAICS code
423450
Employees
368
Ownership type
A

7 citations on file for this inspection.

1910.146 C01

Serious Gravity 10 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
May 12, 2023
Penalty
Initial $15,625 · Current $15,625
29 CFR  1910.146(c)(1): The employer did not evaluate the workplace to determine if any spaces were permit-required confined spaces:  a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 the employer did not evaluate the coil fan room to determine if it was a permit-required confined space.  An employee was working inside the coil fan room to repair a leak and was unable to evacuate when the strainer cap broke off releasing hot steam into the room causing the employee to sustain burns to his body. This condition exposed employees to delayed rescue in the event of an injury and burns.
Recent events (2)
  • — I (S) $15625
  • — Z (S) $15625

1910.146 C02

Deleted Serious Gravity 10 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
Apr 28, 2023
Penalty
Initial $0 · Current $0
29 CFR  1910.146(c)(2): The employer did not inform exposed employees, by posting danger signs or by any other equally effective means, of the existence and location of and the danger posed by permit spaces:   a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 the employer did not inform exposed employees that the that the coil fan  was a permit-required confined space by posting or labeling the existence and location of the danger posed by permit spaces.  This condition exposed employees to delayed rescue in the event of an injury and burns.
Recent events (2)
  • — I (S) $0
  • — Z (S) $0

1910.146 H01

Serious Gravity 10 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
May 12, 2023
Penalty
Initial $15,625 · Current $0 Reduced
29 CFR  1910.146(h)(1):  Duties of authorized entrants. The employer shall ensure that all authorized entrants:  Know the hazards that may be faced during entry, including information on the mode, signs or symptoms, and consequences of the exposure:   a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 the employer did not ensure that all authorized entrants knew the hazards that could be faced during entry, including information on the mode, signs or symptoms and consequences of the exposure: This condition exposed employees to delayed rescue in the event of an injury and burns.
Recent events (2)
  • — I (S) $0
  • — Z (S) $15625

1910.146 I06

Serious Gravity 5 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
May 12, 2023
Penalty
Initial $0 · Current $0
29 CFR  1910.146(i)(6): The employer did not ensure an attendant monitored activities inside and outside a permit confined space to ensure it was safe for entrants to remain in the space and order the authorized entrants to evacuate the permit space immediately under hazardous conditions:   a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 the employer did not ensure that an attendant was monitored activities inside and outside a permit required confined space to ensure it was safe for entrants while working in permit confined spaces. This condition exposed employees to delayed rescue in the event of an injury and burns.
Recent events (2)
  • — I (S) $0
  • — Z (S) $0

1910.146 D03

Serious Gravity 10 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
May 12, 2023
Penalty
Initial $15,625 · Current $15,625
29 CFR  1910.146(d)(3): Under the permit-required confined space program required by 29 CFR 1910.146(c)(4), the employer did not develop and implement the means, procedures, and practices necessary for safe permit space entry operations, including, but not limited to items (d)(3)(i) through (d)(3)(vi) of this paragraph:   a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 the employer did not develop and implement the means, procedures, and practices necessary to ensure safe operations prior to entry inside the coil fan room while conducting maintenance. This condition exposed employees to burns, struck-by injuries, caught in injuries, falls, and delayed rescue.
Recent events (2)
  • — I (S) $15625
  • — Z (S) $15625

1910.146 G01

Deleted Serious Gravity 10 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
Apr 28, 2023
Penalty
Initial $15,625 · Current $0 Reduced
29 CFR  1910.146(g)(1): The employer did not provide training so that all employees whose work was regulated by 29 CFR 1910.146 (permit required confined spaces) acquired the understanding, knowledge, and skills necessary for the safe performance of the duties assigned under 29 CFR 1910.146:    a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 the employer did not provide training to employees to establish knowledge and skills necessary for the safe performance in permit space operations. This condition exposed employees to hazards of entering a permit required confined space and delayed rescue in the event of an incident.
Recent events (2)
  • — I (S) $0
  • — Z (S) $15625

1910.147 C04 I

Serious Gravity 10 1 instance 47 exposed
Issued
Mar 28, 2023
Abate by
May 12, 2023
Penalty
Initial $15,625 · Current $15,625
29 CFR  1910.147(c)(4)(i): Procedures were not developed, documented and utilized for the control of potentially hazardous energy when employees were engaged in activities covered by this section:   a. Carestream Health, Inc., 2000 Howard Smith Ave West, C-40 bldg. Windsor CO 80550: On October 5, 2022 machine specific, step-by-step procedures were not developed, documented and utilized for the control of potentially hazardous energy when employees were engaged in servicing  a broken line in the coil fan room. This condition exposed employees to broken bones and burns.
Recent events (2)
  • — I (S) $15625
  • — Z (S) $15625

View Carestream Health, INC.'s full OSHA safety record →

This record is reproduced from the U.S. Department of Labor Open Data API (OSHA inspection dataset). The original IMIS detail view is available at OSHA's Establishment Search for activity number 346268139.

Look up any company's OSHA accident reports by company, or browse severe injury reports by year, state, and company.